KLOTZ Impaktor FH6价格|型号 _气溶胶发生器原理


KLOTZ Impaktor FH6

Universal air sampler for control of microbiological air quality

  • 商品编号:KLOTZ Impaktor FH6
  • 货  号:KLOTZ Impaktor FH6
  • 品  牌:德国Klotz(德国Klotz公司是长期致力于激光颗粒测量仪器的研究与生产的专业厂家。公司位于德国斯图加特近郊,产品获得多项专利,并赢得德国国家发明大奖。为世界著名企业配套生产(如西门子、MAHLE)激光元器件及传感器。公司具有强有力的生产能力,除具备SMD全自动贴片机、数控加工中心、数控车床等大批先进设备外,还开设了一流的产品检测中心和齐全的检测仪器和设备。Markus Klotz GmbH的公司简介自1990年以来,我们的主要活动领域包括:粒子测量,粒子计数器分析和医学中的诊断技术液体,空气/气体和粒度分析中的颗粒测量,制药和化学工业中的溶液清洁度以及油类制造和开发医用二极管激光辐射装置设计和制造医学中的电气/光学元件和诊断设备在近红外区域刺激后用于检测龋齿的荧光检测Biochip项目合作伙伴,Baden-Württemberg所有设备和相应的软件都是在我们的工厂生产的。)
  • 计量单位:
  • 市场价: ¥0.00
  • 销售价: 面议
购买数量:
  (库存9999+)

  • 收藏此商品

app hook

The air sampler is a measuring instrument for testing the bioburden of the air, which can be used for both the impaction method as well as the filtration method. During the measuring procedure the air is drawn in by a radial fan specifically developed for the FH6 and the flow ist automatically kept constant (impeller anemometer). The electronics control the measurement cycle. The sample volume selected via menu key flows in a defined measuring time through the filter attachment or via the Petri dish with the culture medium. 

In the event of insufficient air flow the device switches off and the display shows Düse zu (nozzle blocked) or Filter verblockt (filter blocked) with the filtration method. In addition, the failure triggers a triple audio signal. In the event of too high air flow rate (device open, no nozzle inserted, filter torn, no filter inserted) the device switches off likewise and the message Schlitz offen (slot open) or Filter defekt (filter defect) is displayed.

A fully charged battery allows at least 90 consecutive measurements of 5 min. each.

Markus Klotz GmbH

Areas of application:

Hospitals, pharmaceutical, cosmetics and foodstuff industry. Here there are stringent requirements for the bioburden of the ambient air and air conditioning. Bioburden emissions of waste sites, sewage plants and irrigation fields are increasingly attracting attention.

Technical Details:

Method: impaction/filtration

Method: impaction/filtration

Volume flow: 100 l/min. impaction method, 30 l/min. filtration method, electronically controlled

Particle materials: (number of germs on the Petri dish or on the gelatin filter manually counted after treatment)

Operating time: 7 h battery operation

Power supply: 16.8 V/charge time 4.5 h / 450 mA

Dimensions (h x w x d): 250 x 150 x 150 mm

Weight / material: 3 kg aluminum / titan